seabreeze

tested on simulator

Tier Tier 2
Protocol Ocean spectrometers through python-seabreeze
Covers Ocean Insight and Ocean Optics USB spectrometers (USB2000+, USB4000, Flame), Ocean HR, QE Pro and Maya spectrometers, Ocean STS and SPARK spectrometers
Hazard default low
Conformance full
Tested with a simulated USB2000+ looking at a halogen lamp through a dye, behind the same Spectrometer methods

Estimated reach. About 30 spectrometer models. python-seabreeze wraps Ocean's open SeaBreeze library, the device layer Ocean ships for its USB spectrometers, and its pure Python backend reimplements the same protocols. Its device table lists about 30 models across the USB, Flame, HR, QE, Maya, STS and SPARK lines, which are some of the most common spectrometers in teaching and research labs. The count is read from the library's device list.

The seabreeze driver reads Ocean Insight and Ocean Optics spectrometers through python-seabreeze 2.11. One device file is one spectrometer, and each scan becomes readings such as the spectrum as a chart, the peak, the counts at a wavelength, and absorbance against a stored dark and reference. describe drafts a file from the spectrometer's wavelengths and integration limits.

The driver is for lab and research spectrometers. OpenHAL has no driver for clinical or patient-connected medical devices, see ADR 0024.

Supports

NameKindMapWhat the driver reads
spectrumread, image{ spectrum: plot }the scan as a PNG chart, wavelength against counts
peak_wavelengthread, length{ peak: wavelength }the wavelength of the highest pixel
peak_intensityread, dimensionless{ peak: intensity }the highest pixel's counts
any nameread, dimensionless{ wavelength: 532 nm }the counts at one wavelength, averaged over 5 pixels
any nameread, dimensionless{ band: [500 nm, 550 nm] }the mean counts over a band
any nameread, dimensionless{ absorbance: 520 nm }-log10 of the sample over the reference, both less the dark
any nameread, percent{ transmittance: 520 nm }the sample over the reference, both less the dark
saturatedread, bool{ status: saturated }true when any pixel reaches the detector's maximum
integration_timewrite, time{ setting: integration_time }integration_time_micros
store_dark, store_reference, clear_referencesaction{ command: store_dark }keep the next scan, or forget both

Behavior

  • Every reading comes from a scan averaged over average scans, and readings within scan_age share one scan.
  • integration_time reads the value last set, or unknown before the first write, since seabreeze reads no integration time back.
  • A new integration time clears the stored dark and reference.
  • For absorbance, block the light and run store_dark, put the blank in and run store_reference, then read the sample. Without both, absorbance and transmittance reads fail with the missing step.
  • The chart colors each wavelength from 380 to 750 nm, and the counts axis runs to the detector maximum, so a saturated scan reaches the top. Numbers come from the reading channels.
  • discover() lists spectrometers from USB descriptors without opening any.

The simulated spectrometer

seabreeze: sim is a simulated USB2000+, 2048 pixels from 340 to 1025 nm with a 16 bit detector, behind the same Spectrometer methods as hardware. A 2900 K halogen lamp shines through a cuvette of dye with one absorbance band at 520 nm. Counts grow with integration time up to 65535, with dark counts, shot noise and read noise. sim:<name> makes a separate simulated spectrometer.

Setup

  1. Install with uv tool install --compile-bytecode "halkit[seabreeze] @ git+https://github.com/sohan-bhat/OpenHAL".
  2. On Linux, install the udev rules with uv run seabreeze_os_setup, and close OceanView, which holds the spectrometer.
  3. Start from absorbance-spectrometer.hal.yaml or describe.
KeyDefaultMeaning
spectrometerautoSerial number, auto for the first found, or sim and sim:<name>
backendcseabreezecseabreeze with built-in USB code, or pyseabreeze, which needs pyusb and libusb
timeout5 sLongest scan, every averaged scan included, at most 9 s
scan_age500 msScan reuse window
average1Scans averaged per reading, 1 to 100
dark_correctionfalseSubtract the detector's electric dark pixels
nonlinearity_correctionfalseApply the stored nonlinearity correction

Known limitations

  • No external triggers, light source or shutter features, or thermoelectric cooler.
  • Integration time times averaged scans must fit timeout.
  • The stored dark and reference are lost on a hub restart.

Hazard defaults

low, with no safe state, since a spectrometer only measures. A light source such as a laser needs a separate device file with a separate hazard.